Configuration 31
Scale Calibration Menu
Calibration can be performed in two places within the menu: the CALIBR menu shown in Figure 3-6 on page 26 is
an in-depth scale setup and calibration.
A “quick access” calibration is shown in Figure 4-1 on page 55.
Figure 3-9. Scale Calibration Menu Layout
Parameter Choices Description
WZERO Allows you to view or change the zero calibration AD count value.
WVAL Allows you to view and edit the test weight value.
WSPAN Allows you to view or change the span calibration AD count value.
WLIN POINT 1-5 Press ENTER to display and edit test weight and calibration values for up to five linearization
points. Perform linear calibration only after WZERO and WSPAN have been set.
Parameters include CAL1 — View or change linear point calibration value.
WGT1 — Sets test value.
REZERO Removes offset from zero and span calibrations. Can be used during a span calibration
process using hooks and chains.
Table 3-8. Scale Calibration Menu Parameters
Display and edit
test weight value
Same as POINT 1
Previous A/D raw
counts are shown
Press calibrate
softkey to calibrate
zero. Press menu
to cancel
calibrating.
Previous A/D raw
counts are shown
Press calibrate
softkey to calibrate
zero. Press menu
to cancel
calibrating.
Calibrating, please wait
Press enter to
remove offset from
zero and span
calibrations
POINT 1
POINT 2 POINT 3
POINT 4
POINT 5
WZERO WVAL WSPAN REZEROWLIN
WGT 1
CAL1
CALIBR