Appendix 97
Where:
<STX> STX character
<P> Polarity
<W7> Seven characters of net data with decimal pt
<U> Units
<M> Mode
<S> Status
<CR> Carriage return
Note that industrial serial scales (IND SC) do not require the <M>, <U> and <S> identifiers. However, the units
and number of decimal places must be specified. Units can be selected from the FORMAT menu; decimal places
should be indicated with a w-spec identifier. For example, a seven-digit weight reading requiring two decimal
places should be specified as <W7.2> rather than <W7>.
To change or match formats, use the same path to change formats.
9.9 Regulatory Mode Functions
The function of the front panel TARE and ZERO keys depends on the value specified for the REGULAT parameter
on the FEATURE menu.
Table 9-5 describes the function of these keys for the NTEP, CANADA, OIML, and
NONE regulatory modes. TARE and ZERO key functions are configurable when the REGULAT mode is set to
INDUST (see
Table 9-6 on page 97).
Table 9-6 lists the subparameters available when configuring a scale using INDUST mode. The table includes the
default values of the INDUST subparameters and the effective (not configurable) values used by the NTEP,
CANADA, OIML, and NONE regulatory modes.
REGULAT
Parameter Value Weight on Scale Tare in System
Front Panel Key Function
TARE ZERO
NTEP zero or negative no no action ZERO
yes CLEAR TARE
positive no TARE
yes TARE
CANADA zero or negative no no action ZERO
yes CLEAR TARE
positive no TARE
yes no action
OIML zero or negative
(not to exceed -20dd)
no no action ZERO
yes CLEAR TARE ZERO and CLEAR TARE
positive no TARE ZERO
yes TARE ZERO and CLEAR TARE if weight is
within ZRANGE.
No action if weight is outside of
ZRANGE
NONE zero or negative no TARE ZERO
yes CLEAR TARE
positive no TARE
yes CLEAR TARE
Table 9-5. TARE and ZERO Key Functions for REGULAT Parameter Settings
REGULAT / INDUST Parameter REGULAT Mode
Parameter Name Tex t P ro m p t INDUST NTEP CANADA OIML NONE
SNPSHOT Display or Scale weight source DISPLAY DISPLAY DISPLAY DISPLAY SCALE
ZTARE Remove tare on ZERO NO NO NO YES NO
KTARE Always allow keyed tare YES YES NO YES YES
MTARE Multiple tare action REPLACE REPLACE NOTHING REPLACE REMOVE
NTARE Allow negative tare NO NO NO NO YES
Table 9-6. REGULAT / INDUST Mode Parameters, Comparison with Effective Values of Other Modes