Configuration 15
3.0 Configuration
To configure the 820i indicator, the indicator must be
placed in setup mode. The setup switch is accessed by
removing the large fillister head screw on the bottom
of the enclosure. Switch position is changed by
inserting a screwdriver into the access hole and
pressing the switch.
When the indicator is placed in setup mode, a series of
menus
is shown across the top of the display, along
with the words
Scale Configuration. The SCALES menu
is highlighted as the first used to configure the
indicator. Detailed descriptions of these menus are
provided in Section 3.2.
When configuration is complete, press the Exit or Save
and Exit
softkey to exit setup mode, then replace the
setup switch access screw.
• The
Exit softkey exits setup mode without saving
parameter changes to NV RAM. Changes made
to the configuration remain in the system until
indicator power is cycled.
•
Save and Exit writes all parameter changes to NV
RAM before returning to normal mode.
3.1 Configuration Methods
The 820i indicator can be configured by using the
front panel keys to navigate through a series of
configuration menus or by sending commands or
configuration data to an indicator serial port.
Configuration using the menus is described in
Section 3.1.3.
Configuration using the serial port can be
accomplished
using either the serial command set
described in Section 9.0 on page 79 or the
Revolution® configuration utility.
Note
Some configuration parameters, such as
those used to configure the
820i display,
cannot be accessed through the
configuration menus.
Revolution provides the most
complete and efficient configuration interface for the
820i.
3.1.1 Revolution Configuration
The Revolution configuration utility provides the
preferred method for configuring the
820i indicator.
Revolution runs on a personal computer to set
configuration parameters for the indicator. When
Revolution configuration is complete, configuration
data is downloaded to the indicator.
Revolution supports both uploading and downloading
of indicator configuration data. This capability allows
configuration data to be retrieved from one indicator,
edited, then downloaded to another indicator with an
identical hardware configuration.
To use
Revolution, do the following:
1. Install
Revolution on an IBM-compatible
personal computer. See Section 5.0 on
page 51 for detailed hardware and software
requirements.
2. With both indicator and PC powered off,
connect the PC serial port to the RS-232 pins
on
the indicator serial port or to the optional
USB communications card.
3. Power up the PC and the indicator. Use the
setup
switch to place the indicator in setup
mode.
4. Start the
Revolution program.
Figure 3-1. Revolution Display
Revolution provides online help for each of its
configuration displays. Parameter descriptions
provided in this manual for front panel configuration
can also be used when configuring the indicator using
Revolution: The interface is different, but the
parameters set are the same.
See Section 5.0 on page 51 for more information about
using
Revolution to configure the 820i.
3.1.2 Serial Command Configuration
The serial command set can be used to configure the
820i indicator using either a personal computer,
terminal, or remote keyboard. Like
Revolution, serial
command configuration sends commands to the
indicator serial port; unlike
Revolution, serial
commands can be sent using any external device
capable of sending ASCII characters over a serial
connection.