Rice Lake 340-10 Series Operations Instructions

Browse online or download Operations Instructions for Kitchen scales Rice Lake 340-10 Series. Rice Lake 340-10 Series Operating instructions [en] User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - Instructions

118580Rice Lake Digital Wheelchair Scale340-10 SeriesSoftware Revision 11387Technical and Operating InstructionsTo be the best by every measure

Page 2

6 Wheelchair Scale Technical and Operating Instructions2.5 Hooking up the AC Power ConnectionsThe Digital Wheelchair Scale has a 120 VAC adaptor or 2

Page 3 - Contents

Wheelchair Scale Technical and Operating Instructions - Scale Assembly 72.6 Getting Ready to Weigh a PatientOnce the scale is properly unpacked and s

Page 4

8 Wheelchair Scale Technical and Operating Instructions3.0 Scale OperationThe display has various front panel keys. They are shown below and their fu

Page 5 - I ntern et

Wheelchair Scale Technical and Operating Instructions - Scale Operation 93.2 Using the Body Mass Index (BMI) FunctionBody mass index (BMI) is the rel

Page 6 - 2.0 Scale Assembly

10 Wheelchair Scale Technical and Operating Instructions3.3 Using the Tare FunctionYou can use the tare function for deducting an extra weight (such

Page 7 - 2.3 Setting Up Your Scale

Wheelchair Scale Technical and Operating Instructions - Scale Configuration 114.0 Scale ConfigurationOptions and parameter setup are done through the

Page 8

12 Wheelchair Scale Technical and Operating InstructionsVarious parameters can be set up while in programming mode. Figure 4-2. Programming Mode Menu

Page 9

Wheelchair Scale Technical and Operating Instructions - Scale Configuration 13The following table lists the various display messages and sequence whe

Page 10 - Battery Charging

14 Wheelchair Scale Technical and Operating Instructions13 Message style on weight algorithm MESS <-> WEIGH This displays the message that will

Page 11

Wheelchair Scale Technical and Operating Instructions - Scale Configuration 15Option 6 allows you to either enable or disable the hold function on th

Page 13

16 Wheelchair Scale Technical and Operating Instructions4.4 Reset To Factory DefaultsThe configuration parameters may be reset to factory defaults wh

Page 14 - 3.3 Using the Tare Function

Wheelchair Scale Technical and Operating Instructions - Scale Calibration 175.0 Scale CalibrationBefore you can calibrate the scale, verify and set a

Page 15 - 4.0 Scale Configuration

18 Wheelchair Scale Technical and Operating Instructions6.0 RS-232 CommunicationThe scale comes with an RS-232 port which enables weight data to be t

Page 16 - 4.3 Numeric Data Entry

Wheelchair Scale Technical and Operating Instructions - RS-232 Communication 196.3 ESC Protocol (configuration option #9 set to 0)The ESC Protocol di

Page 17 - *With the scale turned off

20 Wheelchair Scale Technical and Operating Instructions3. Double click on the Rswin.exe file and the following screen appears.Figure 6-1. Rswin Scre

Page 18

Wheelchair Scale Technical and Operating Instructions - RS-232 Communication 21Examples of what you would see in the HEX screen are shown in Figure 6

Page 19

22 Wheelchair Scale Technical and Operating Instructions6.4 Sample and Explanation of ESC ProtocolWhen the scale measures weight and sends this over

Page 20 - 4.4 Reset To Factory Defaults

Wheelchair Scale Technical and Operating Instructions - RS-232 Communication 23If the battery is okay, the scale will reply with the following value:

Page 21 - 5.0 Scale Calibration

24 Wheelchair Scale Technical and Operating Instructions7.0 Troubleshooting and TestingRefer to the following instructions to check and correct any f

Page 22 - 6.0 RS-232 Communication

Wheelchair Scale Technical and Operating Instructions - Troubleshooting and Testing 257.1 Test ModeThe test mode menu is a special mode used for chec

Page 23

Contents1.0 Introduction...

Page 24 - Figure 6-1. Rswin Screen

26 Wheelchair Scale Technical and Operating Instructions5Checks for current battery levelbAt <-> XXX or nO bAt If the nO bAt is displayed, ther

Page 25

Wheelchair Scale Technical and Operating Instructions - Maintenance 278.0 MaintenanceThe following section provides instructions for maintaining and

Page 26 - &4$3&4$&

28 Wheelchair Scale Technical and Operating Instructions8.3 Trimming ProcedureTrimming is the process of equalizing the output from multiple individu

Page 27 - DIRECTIONOFCOMMUNICATION

Wheelchair Scale Technical and Operating Instructions - Maintenance 293. Replace the same test load over each cell in turn measuring load cell one (1

Page 28

30 Wheelchair Scale Technical and Operating Instructions9.0 Wheelchair Scale SpecificationsPower120 VAC-9VDC-50Hz / 230 VAC-9VDC-50HzBattery TypeSeal

Page 29 - Calibr. Jumper

Wheelchair Scale Technical and Operating Instructions - Wheelchair Scale Specifications 31For More InformationSystem Manuals• Digital Wheelchair Scal

Page 30 - Table 7-2. Test Mode Menu

32 Wheelchair Scale Technical and Operating InstructionsDigital Wheelchair Scale Limited WarrantyRice Lake Weighing Systems (RLWS) warrants that all

Page 33

ii Wheelchair Scale Technical and Operating InstructionsRice Lake continually offers web-based video training on a growing selection of product-relat

Page 35 - For More Information

2 Wheelchair Scale Technical and Operating Instructions2.0 Scale Assembly2.1 Unpacking Your ScalePlace the unopened box in an open area that has ampl

Page 36

Wheelchair Scale Technical and Operating Instructions - Scale Assembly 32.3 Setting Up Your ScaleUse the following steps to set up the Digital Wheelc

Page 37

4 Wheelchair Scale Technical and Operating Instructions2.4 Attaching the Feet to the ScaleThe Digital Wheelchair Scale comes with four feet that must

Page 38 - PN 118580 01/12

Wheelchair Scale Technical and Operating Instructions - Scale Assembly 55. Loosen the handles (shown in Figure 2-6) and set upright as shown in Figur

Comments to this Manuals