106992Summit SB-1000 Bariatric Wheelchair ScaleInstallation & Operation Manual®
6 Summit SB-1000 Bariatric Wheelchair Scale Installation ManualWith all potentiometers at full signal, place the test weight over one corner and reco
Indicator Installation and Configuration 74.0 Indicator Installation and Configuration4.1 Indicator InstallationThe Summit SB-1000 Bariatric Wheelchai
8 Summit SB-1000 Bariatric Wheelchair Scale Installation ManualLoad Cell ConnectionsTo gain access to the load cell connection point, remove the four
Indicator Installation and Configuration 94.3.1 Configuration ModeTo get into the configuration mode, turn the scale off and remove the configuration
10 Summit SB-1000 Bariatric Wheelchair Scale Installation ManualThe following table lists the various display messages and sequence when setting up t
Indicator Installation and Configuration 1112 Message style on weight algorithmMESS <-> WEIGH This displays the message that will show on the in
12 Summit SB-1000 Bariatric Wheelchair Scale Installation ManualOption 5 allows for live or dynamic weighingOP5 = 0 Hold the Kg-Lb key for browsing t
Indicator Installation and Configuration 134.4 Reset To Factory DefaultsThe configuration parameters may be reset to factory defaults while in the con
14 Summit SB-1000 Bariatric Wheelchair Scale Installation Manual5.0 Scale CalibrationBefore you can calibrate the scale, verify and set all scale par
Scale Operation 156.0 Scale OperationThe scales have the capability of performing different operations beyond just calculating weight. The various ope
16 Summit SB-1000 Bariatric Wheelchair Scale Installation Manualincrements. Press the BMI key again to accept the inches value. The final height valu
RS-232 Communications 177.0 RS-232 CommunicationsThe scale comes with an RS-232 port which enables weight data to be transmitted to other equipment, s
18 Summit SB-1000 Bariatric Wheelchair Scale Installation ManualAn Escape Protocol is where the escape <ESC> is used to indicate that there is
RS-232 Communications 193. Double click on the Rswin.exe file and the following screen appears.Figure 7-1. Rswin Main Screen4. Click on FILES and in t
20 Summit SB-1000 Bariatric Wheelchair Scale Installation ManualExamples of what you would see in the HEX screen are shown in Figure 7-3.Figure 7-3.
RS-232 Communications 21When the user wants to diagnose any problems with the scale, the operator will have to ask the scale to send the error data (i
22 Summit SB-1000 Bariatric Wheelchair Scale Installation Manual8.0 Service InformationThe following sections describe basic service and maintenance
Service Information 238.2 Test ModeThe test mode menu is a special mode used for checking four very important parameters which are useful in knowing t
24 Summit SB-1000 Bariatric Wheelchair Scale Installation ManualTable 8-2 lists the various display messages when testing the scale.NOTE: <-> m
Service Information 258.3 Load Cell ReplacementReplacement load cells can be ordered from Rice Lake Weighing Systems. To replace a load cell, lift the
iContents1.0 Introduction and System Overview... 12.0 Sca
26 Summit SB-1000 Bariatric Wheelchair Scale Installation ManualFigure 8-4. Foot Pad - Side ViewRoute the load cell cables near each corner so that t
Bariatric Wheelchair Scale Specifications 279.0 Bariatric Wheelchair Scale SpecificationsElectrical GroundingFor systems where the scale is connected
28 Summit SB-1000 Bariatric Wheelchair Scale Installation ManualSummit SB-1000 Limited WarrantyRice Lake Weighing Systems (RLWS) warrants that all RL
ii Summit SB-1000 Bariatric Wheelchair Scale Installation Manual
Introduction and System Overview 11.0 Introduction and System OverviewThe Summit® SB-1000 Bariatric Wheelchair Scale is a heavy-duty scale that easily
2 Summit SB-1000 Bariatric Wheelchair Scale Installation Manual2.0 Scale InstallationThe following sections describe the correct installation procedu
Scale Installation 32.4 Bariatric Wheelchair Scale Feet Assembly and AdjustmentThe scale feet are shipped attached to the scale. If the feet have to b
4 Summit SB-1000 Bariatric Wheelchair Scale Installation Manualservicing or cleaning.Figure 2-3. Load Cell Protection8. When the interface cable is p
Adjustments and Calibration 53.0 Adjustments and CalibrationThe following sections describe adjustments that need to be made to the Summit SB-1000 Bar