24 RL-TC Series Operation Manual
8.2 Setup of Limit Values by Actual Quantity Loads
1 Starting the limit function
Press and hold down the Target key.
Release the key when [L. SEt] is
displayed.
The currently set lower-limit value
flashes.
2 Loading the sample for the lower-limit value
Load the sample of the lower-limit value
on the measurement pan.
3 Saving the lower-limit value
Press the Memory key.
After the lower-limit value has been
saved, the balance
displays it briefly and
proceeds to the following setup.
If One-point setup was chosen, the
setup is complete.
The display now changes to
[
H. SET], indicating that the upper-
limit
value can be set.
The currently set upper-limit value
flashes.
4 The upper-limit
value setup
5 Loading the sample of the upper-limit value
Load the sample of the upper-
limit value
on the measurement pan.
6 Saving the upper-limit value
Press the Memory key.
After saving the upper-limit value, the
balance displays it briefly and terminates
the setup.